TY - JOUR TI - Field Evaluation of the New Rapid NG-Test® SARS-CoV-2 Ag for Diagnosis of COVID-19 in the Emergency Department of an Academic Referral Hospital AU - Avgoulea, K. AU - Beredaki, M.-I. AU - Vourli, S. AU - Siopi, M. AU - Siafakas, N. AU - Pournaras, S. JO - Frontiers in public health PY - 2022 VL - 10 TODO - null SP - null PB - Frontiers Media S.A SN - 2296-2565 TODO - 10.3389/fpubh.2022.840984 TODO - virus antigen, adult; diagnosis; emergency; hospital; hospital emergency service; human; pandemic; patient referral; sensitivity and specificity, Adult; Antigens, Viral; COVID-19; Emergencies; Emergency Service, Hospital; Hospitals; Humans; Pandemics; Referral and Consultation; SARS-CoV-2; Sensitivity and Specificity TODO - Background: As the COVID-19 pandemic resurges affecting large numbers of patients, rapid, and accurate diagnosis using point-of-care tests is very important. Objectives: To evaluate the NG-Test® SARS-CoV-2 Ag (NG-Test) immunoassay for qualitative detection of SARS-CoV-2 antigen in nasopharyngeal (NP) and oropharyngeal (OP) samples compared with RT-PCR, in patients attending the Emergencies of an academic referral hospital. Methods: All adult ambulatory patients presenting to the Emergencies of “Attikon” University hospital (Athens, Greece) within three consecutive hours per day between December 2020 and March 2021 and for whom SARS-CoV-2 PCR testing was requested were included. Two NP and one OP samples obtained from each participant were analyzed to determine the diagnostic performance [sensitivity, specificity, positive/negative predictive values (PPV/NPV)] of the NG-Test (NP/OP swabs) in comparison to the reference RT-PCR (NP swab). Results: Overall, 134/263 (51%) patients tested were RT-PCR positive, whereof 108 (overall sensitivity 81%, 95% CI 73–87%) were NP NG-Test positive (PPV 99%, NPV 83%) and 68 (overall sensitivity 51%, 95% CI 42–59%) were OP NG-Test positive (PPV 100%, NPV 66%). The test's specificity (95% CI) was 99% (95–100%) and 100% (96–100%) for NP and OP swabs, respectively. The assay's sensitivity (95% CI) for high viral load (Ct ≤25) was 99% (92–100%) and 71% (60–81%) for NP and OP swabs, respectively. Conclusions: NG-Test using NP swabs detected almost all patients with high viral loads, showing satisfactory performance as a point-of-care test for NP samples obtained from patients with acute infection. Copyright © 2022 Avgoulea, Beredaki, Vourli, Siopi, Siafakas and Pournaras. ER -