Comparison of Goldmann applanation tonometry and Pascal Contour Tonomtery in relatopn to central corneal thickness and corneal curvature

Doctoral Dissertation uoadl:1326914 347 Read counter

Unit:
Τομέας Χειρουργικής
Library of the School of Health Sciences
Deposit date:
2017-01-20
Year:
2016
Author:
Andreanos Konstantinos
Dissertation committee:
Mόσχου Ευαγγελία Μαρία Αναπληρώτρια Καθηγήτρια, Ιατρική σχολή, ΕΚΠΑ
Κουτσανδρέα Χρυσάνθη, Καθηγήτρια, Ιατρική σχολή, ΕΚΠΑ
Παπακωνσταντίνου Δημήτριος, Αναπληρωτής Καθηγητής, Ιατρική σχολή, ΕΚΠΑ
Λαδάς Ιωάννης, Καθηγητής, Ιατρική σχολή, ΕΚΠΑ
Κυμιωνής Γεώργιος Αναπληρωτής Καθηγητής, Ιατρική σχολή, ΕΚΠΑ
Μπρούζας Δημήτριος, Αναπληρωτής Καθηγητής, Ιατρική σχολή, ΕΚΠΑ
Δρούτσας Κωνσταντίνος Λέκτορας, Ιατρική σχολή, ΕΚΠΑ
Original Title:
Μελέτη της ενδοφθάλμιας πίεσης με την Dynamic Contour Tonometry και την Goldmann Applanation Tonometry σε σχέση με το πάχος και την καμπυλότητα του κερατοειδούς.
Languages:
Greek
Translated title:
Comparison of Goldmann applanation tonometry and Pascal Contour Tonomtery in relatopn to central corneal thickness and corneal curvature
Summary:
Purpose : To assess the influence of central corneal thickness and corneal curvature in tonometry measurements taken by Goldmann applanation tonometry (GAT) and Pascal Dynamic Contour Tonometry (DCT).
Methods: Prospective study of 185 eyes from 97 subjects, attending outpatient ophthalmology appointments, who underwent intraocular pressure measurements by Goldmann applanation tonometer (GAT) and Pascal Dynamic contour tonometer (DCT). Central corneal thickness (cct) and corneal curvature were obtained using ultrasound pachymetry and orbscan topography respectively. All measurements carried out among males and females during the period 2009 to 2012. Apart from the usual descriptive and exploratory data analysis, linear correlation coefficients were estimated, multiple scatter plots were produced and one-way analysis of variance was performed.
Results: Mean IOP measurements obtained were 17,21 ± 4,10mmHg by DCT and 13,23 ± 4,07 mmHg by GAT. Mean difference between the GAT and DCT measurements was 3,88 ±2,8 mmHg. Mean CCT and Corneal curvature were 522,78 ± 52 μm and 43,83 ± 2,9823 Diopters respectively.
Conclusions: Intraocular pressure measured by GAT was consistently lower when compared with DCT, and this difference was greatest with thinner CCT. Flat corneas seem to influence GAT readings compared to DCT.
Keywords:
Glaucoma, Tonometry , Corneal Thickness, Corneal curvature
Index:
No
Number of index pages:
0
Contains images:
Yes
Number of references:
88
Number of pages:
161
File:
File access is restricted only to the intranet of UoA.

ΔΙΑΤΡΙΒΗ.pdf
2 MB
File access is restricted only to the intranet of UoA.