Comparative analysis of mechanical properties of orthodontic aligners produced by different contemporary 3D printers

Επιστημονική δημοσίευση - Άρθρο Περιοδικού uoadl:2989887 22 Αναγνώσεις

Μονάδα:
Ερευνητικό υλικό ΕΚΠΑ
Τίτλος:
Comparative analysis of mechanical properties of orthodontic aligners produced by different contemporary 3D printers
Γλώσσες Τεκμηρίου:
Αγγλικά
Περίληψη:
Objective: The aim of this study was to compare the mechanical properties of orthodontic aligners among different commercially available 3D printing devices. Materials and Methods: Five 3D printers (Ka:rv LP 550, Swinwon; “KAR”), (L120, Dazz 3D; “L12”), (MiiCraft 125, Miicraft Jena; “MIC”), (Slash 2, Uniz; “SLS”) and (Pro 95, SprintRay; “PRO”) were used to prepare orthodontic aligners with dental resin (Tera Harz TC-85DAW, Graphy). The central incisors of each aligner were cut, prepared and evaluated in terms of Martens-Hardness (HM), indentation-modulus (EIT) and elastic-index (ηIT) as per ISO14577-1:2002. Force-indentation curves were recorded and differences among printers were checked with generalized linear regressions (alpha=5%). Results: Statistically significant differences were seen for all mechanical properties (P <.05), which were in descending order: HM (N/mm2) as median (Interquartile Range [IQR]): SLS 108.5 (106.0-112.0), L12 103.0 (102.0-107.0), KAR 101.5 (97.5-103.0), MIC 100.0 (97.5-101.5) and PRO 94.0 (93.0-96.0); EIT (MPa) as mean (Standard Deviation [SD]): SLS 2696.3 (124.7), L12 2627.8 (73.5), MIC 2566.2 (125.1), KAR 2565.0 (130.2) and PRO 2491.2 (53.3); and ηIT (%) as median (IQR): SLS 32.8 (32.3-33.1), L12 31.6 (30.8-32.3), KAR 31.3 (30.9-31.9), MIC 30.5 (29.9-31.2) and PRO 29.5 (29.1-30.0). Additionally, significant differences existed between liquid crystal display (LCD) and digital light processing (DLP) printers for HM (P <.001), EIT (P =.002) and ηIT (P <.001), with aligners from the former having higher values than aligners from the latter printer. Conclusion: Under the limitations of this study, it may be concluded that the mechanical properties of 3D-printed orthodontic aligners are dependent on the 3D printer used, and thus, differences in their clinical efficacy are anticipated. © 2021 The Authors. Orthodontics & Craniofacial Research published by John Wiley & Sons Ltd.
Έτος δημοσίευσης:
2021
Συγγραφείς:
Zinelis, S.
Panayi, N.
Polychronis, G.
Papageorgiou, S.N.
Eliades, T.
Περιοδικό:
Orthodontics and Craniofacial Research
Εκδότης:
John Wiley and Sons Inc
Επίσημο URL (Εκδότης):
DOI:
10.1111/ocr.12537
Το ψηφιακό υλικό του τεκμηρίου δεν είναι διαθέσιμο.