Abstract:
The development of new hydrogen selective neutral carriers based on 3-hydroxypicolinic acid derivatives, and its application to pH selective liquid membrane ISE based on these compounds is described. The optimization of the sensor has been based on the use of different plasticizers, additives, and mole ratios of the picolinate derivatives. Based on these results it is shown that the membrane composition can be adjusted for the measurement of pH at very low pH values (down to pH -0.5) and up to at least pH 10. The ability of the presented carriers to measure hydrogen activities up to 3N gives new directions in the use of these sensors for a variety of critical applications.
Publication year:
2002
Authors:
Maria Fouskaki
Thanasis Gimisis
Nikolas0.25emA. Chaniotakis
Journal:
Electroanalysis
Publisher:
Wiley
Volume:
14
Number:
9
Pages:
593
Keywords:
PH selective,[3-Hydroxypicolinic acid derivatives
Main subject category:
Science
DOI:
10.1002/1521-4109(200205)14:9<593::aid-elan593>3.0.co;2-8
The digital material of the item is not available.