Validation of an Automatic Wrist Blood Pressure Monitor in a Critical Care Setting
Γλώσσες Τεκμηρίου:
Αγγλικά
Περίληψη:
Invasive blood pressure (IBP) is the gold standard method for BP monitoring in critically ill patients. However, due to practical difficulties and complications, noninvasive methods may offer an alternative. We aim to evaluate an automatic oscillometry-based wrist BP monitor in critically ill patients compared to IBP. Forty five adult patients were included, and systolic, diastolic and mean BP (SBP, DBP, MBP) were simultaneously measured by IBP in the radial artery and by non-invasive blood pressure (NIBP) monitoring using the OMRON RS7 automatic wrist monitor. NIBP and IBP (mean ± SD) were: SBP 114 ± 18 vs 130 ± 18; DBP 63 ± 13 vs 63 ± 13; MBP 80 ± 13 vs 86 ± 15 mmHg. Bland-Altman analysis of the differences (NIBP-IBP) mean bias was: SBP-16; DPB 0.5 and MBP-6. The OMRON wrist monitor may not be sufficiently accurate compared to IBP in critically ill patients. © 2020 The authors and IOS Press.
Έτος δημοσίευσης:
2020
Συγγραφείς:
Karampela, I.
Dalamaga, M.
Stratigou, T.
Antonakos, G.
Nikolopoulos, M.
Tzortzis, E.
Diomidous, M.
Περιοδικό:
Studies in Health Technology and Informatics
Εκδότης:
IOS Press BV
Τόμος:
272
Σελίδες:
171-174
Λέξεις-κλειδιά:
Blood; Blood pressure; Medical informatics; Noninvasive medical procedures, Adult patients; Blood pressure monitors; Critical care; Critically-ill patients; Gold standards; Noninvasive methods; Oscillometry; Radial artery, Mercury compounds, blood pressure; blood pressure measurement; blood pressure monitor; human; intensive care; oscillometry; wrist, Blood Pressure; Blood Pressure Determination; Blood Pressure Monitors; Critical Care; Humans; Oscillometry; Wrist
Επίσημο URL (Εκδότης):
DOI:
10.3233/SHTI200521
Το ψηφιακό υλικό του τεκμηρίου δεν είναι διαθέσιμο.