On the use of a novel Ferrous Xylenol-orange gelatin dosimeter for HDR brachytherapy commissioning and quality assurance testing

Επιστημονική δημοσίευση - Άρθρο Περιοδικού uoadl:3072322 6 Αναγνώσεις

Μονάδα:
Ερευνητικό υλικό ΕΚΠΑ
Τίτλος:
On the use of a novel Ferrous Xylenol-orange gelatin dosimeter for HDR brachytherapy commissioning and quality assurance testing
Γλώσσες Τεκμηρίου:
Αγγλικά
Περίληψη:
Purpose: To evaluate a commercially available Ferrous-Xylenol Orange-Gel (FXG) dosimeter (TrueView™) coupled with Optical-Computed Tomography (OCT) read out, for 3D dose verification in an Ir-192 superficial brachytherapy application. Methods: Two identical polyethylene containers filled with gel from the same batch were used. One was irradiated with an 18 MeV electron field to examine the dose-response linearity and obtain a calibration curve. A flap surface applicator was attached to the other to simulate treatment of a skin lesion. The dose distribution in the experimental set up was calculated with the TG-43 and the model based dose calculation (MBCA) algorithms of a commercial treatment planning system (TPS), as well as Monte Carlo (MC) simulation using the MCNP code. Measured and calculated dose distributions were spatially registered and compared. Results: Apart from a region close to the container's neck, where gel measurements exhibited an over-response relative to MC calculations (probably due to stray light perturbation), an excellent agreement was observed between measurements and simulations. More than 97% of points within the 10% isodose line (80 cGy) met the gamma index criteria established from uncertainty analysis (5%/2 mm). The corresponding passing rates for the comparison of experiment to calculations using the TG-43 and MBDCA options of the TPS were 57% and 92%, respectively. Conclusion: TrueView™ is suitable for the quality assurance of demanding radiotherapy applications. Experimental results of this work confirm the advantage of the studied MBDCA over TG-43, expected from the improved account of scatter radiation in the treatment geometry. © 2017 Associazione Italiana di Fisica Medica
Έτος δημοσίευσης:
2018
Συγγραφείς:
Pappas, E.P.
Peppa, V.
Hourdakis, C.J.
Karaiskos, P.
Papagiannis, P.
Περιοδικό:
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS
Εκδότης:
Associazione Italiana di Fisica Medica
Τόμος:
45
Σελίδες:
162-169
Λέξεις-κλειδιά:
polyethylene; gelatin; iridium; phenol derivative; sulfoxide; xylenol orange, algorithm; Article; brachytherapy; calibration; computer assisted tomography; controlled study; dose response; dosimetry; ferrous xylenol orange gel dosimeter; flap surface applicator; gel; irradiation; model based dose calculation algorithm; Monte Carlo method; optical computed tomography; radiation dose distribution; TG 43 based algorithm; brachytherapy; computer simulation; devices; dosimeter; electron; health care quality; human; optical tomography; procedures; radiation response; radiometry; uncertainty, Algorithms; Brachytherapy; Calibration; Computer Simulation; Dose-Response Relationship, Radiation; Electrons; Gelatin; Humans; Iridium Radioisotopes; Monte Carlo Method; Phenols; Polyethylene; Quality Assurance, Health Care; Radiation Dosimeters; Radiometry; Sulfoxides; Tomography, Optical; Uncertainty
Επίσημο URL (Εκδότης):
DOI:
10.1016/j.ejmp.2017.12.015