Radiological evaluation of the metal-bone interface of a porous tantalum monoblock acetabular component

Επιστημονική δημοσίευση - Άρθρο Περιοδικού uoadl:3098989 14 Αναγνώσεις

Μονάδα:
Ερευνητικό υλικό ΕΚΠΑ
Τίτλος:
Radiological evaluation of the metal-bone interface of a porous tantalum
monoblock acetabular component
Γλώσσες Τεκμηρίου:
Αγγλικά
Περίληψη:
Between January 1998 and December 1998, 82 consecutive patients (86
hips) underwent total hip arthroplasty using a trabecular metal
monoblock acetabular component. All patients had a clinical and
radiological follow-up evaluation at six, 12 and 24 weeks, 12 months,
and then annually thereafter. On the initial post-operative radiograph
25 hips had a gap between the outer surface of the component and the
acetabular host bed which from 1 to 5 mm. All patients were followed up
clinically and radiologically for a mean of 7.3 years (7 to 7.5). The 25
hips with the 1 to 5 mm gaps were studied for component at two years
using the Einzel-Bild-Roentgen-Analyse (EBRA) digital measurement At 24
weeks all the post-operative gaps were filled with bone and no
acetabular had migrated. The radiographic outcome of all 86 components
showed no radiolucent lines and no evidence of lysis. No acetabular
implant was revised. There were no dislocations or other complications.
The bridging of the interface gaps (up to 5 mm) by the trabecular metal
monoblock acetabular component indicates the strong osteoconductive, and
possibly osteoinductive, properties of trabecular metal.
Έτος δημοσίευσης:
2006
Συγγραφείς:
Macheras, GA
Papagelopoulos, PJ
Kateros, K
Kostakos, AT and
Baltas, D
Karachalios, TS
Περιοδικό:
JOURNAL OF BONE AND JOINT SURGERY. BRITISH VOLUME
Εκδότης:
BRITISH EDITORIAL SOC BONE & JOINT SURGERY
Τόμος:
88B
Αριθμός / τεύχος:
3
Σελίδες:
304-309
Επίσημο URL (Εκδότης):
DOI:
10.1302/0301-620X.88B3.16940