Τίτλος:
Comparison of three current viral load assays for the quantitation of
human immunodeficiency virus type 1 RNA in plasma
Γλώσσες Τεκμηρίου:
Αγγλικά
Περίληψη:
The LCx HIV RNA quantitative assay (Abbott Laboratories, Delkenheim,
Germany) was compared with the Versant HIV-1 RNA 3.0 (bDNA) assay
(Bayer, Tarrytown, NY) and the COBAS Amplicor HIV-1 Monitor v1.5 test
(Roche Diagnostics, Branchburg, NJ), using plasma samples of various
viral load levels from HIV-1-infected patients. Considering the lower
limit of the linear range of 50 copies/ml of both assays, the detection
range of the LCx was 127/151 (84.1%) versus the 131/151 (86.8%) of the
bDNA 3.0 assay, while overall agreement between the two assays was
93.4% (141/151). LCx and bDNA 3.0 results were found to be strongly
correlated (r = 0.96). The fitted regression line was described by the
equation log(10)(LCx copies/ml) = 0.05 + 1.06 x log(10)(bDNA
3.Ocopies/ml) with 95% CI for the estimated slope and intercept at
1.01, 1.12 and -0.16. 0.26, respectively. Similarly, the detection range
of the LCx was 115/148 (77.7%) versus the 128/148 (86.5%) of the
Monitor v 1.5 test. A 91.2% concordance (135/148) was observed between
these two assays at a cut-off of 50 copies/ml. LCx and Monitor v 1.5
results were highly correlated (r = 0.96). The fitted regression line
was described by the equation log(10)(LCx copies/ml)=0.06 + 1.03 x
log(10)(Monitor v1.5 copies/ml) with 95% Cl for the estimated slope and
intercept at 0.97, 1.09 and -0.16, 0.28, respectively. (C) 2004 Elsevier
B.V. All rights reserved.
Συγγραφείς:
Katsoulidou, A
Papachristou, E
Petrodaskalaki, M
Sypsa, V
and Anastassopoulou, CG
Gargalianos, P
Karafoulidou, A and
Lazanas, M
Kordossis, T
Andoniadou, A
Hatzakis, A
Περιοδικό:
Journal of Virological Methods
Εκδότης:
ELSEVIER SCIENCE BV
Λέξεις-κλειδιά:
HIV-1 RNA; viral load; LCx; branched DNA; COBAS Amplicor Monitor v1.5
DOI:
10.1016/j.jviromet.2004.06.007